Pular para o conteúdo principal
GE Healthcare · MRExcite

Unable To Film due to DASM Not Installed in Software After DASM Replacement.

Symptom

Unable To Film

Error message

scsi_open: /dev/dasm1 open failed - errno 2 Could not initialize_scsi status = ffffffff

dasm_load zero image detected, iid 2 See Long Description Following error logged in /export/home/sdc/logfiles/prslog: dasm_load: zero image detected, iid 2

The printer interface was unable to open the SCSI device driver. Please resume the print job in the Film Queue. See Long Description Following error is logged to /export/home/sdc/logfiles/prslog: MESSAGE from Process 9643>> Fri May 28 11:07:39 2010 (prsjob.c:436) [PRSserver]> The printer interface was unable to open the SCSI device driver. Please resume the print job in the Film Queue.