No Longitudinal Drive Functions, Cradle will not Drive in or out Due to the Table Up micro switch.
Symptom
No Longitudinal Drive Functions, Cradle will not Drive in or out. Also appears none of the buttons work on either side on the front of the magnet enclosure. See long description. The system is a 1.5T Excite which was just upgraded from 11.0 M3 to 11.0 M4. After the upgrade, there are no longitudinal drive functions. The cradle will not drive in or out. It appears that none of the other buttons on the magnet enclosure function as well, on either side. The FE ran the “Magnet Enclosure Button Test” found under Functional Checks in the Service Docs. Looks like none of the buttons are seen. Couple of notes; there are no errors in the log and the TPS reset is successful. Also, if they lower the table, the cradle will retract as normal. So it would appear that there is some longitudinal function, just not with the button pads. Replaced the SRI2 twice and no difference. Also disconnected one magnet enclosure button pad at a time to isolate a bad button pad but no difference.
Error message
2224666 Auto Prescan failed The table is not at scan plane Wed Oct 19 18:25:22 2005 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, The table is not at scan plane.
Applies to
- Models
- Signa, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-136334P56 BASIC SWITCH W/ OT PLUNGER Table Up Limit 46-136334P56 2 SWITCH 1 SPDT, BASIC SWITCH W/ OT PLUNGER.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.