GE Healthcare · MR22280102228016
Array processor AP errors during scanning.
Symptom
Array processor AP errors during scanning. Scans stop (abort) with an AP error logged. It appears that AP crashes are not related to heavy duty recon jobs or high data acquisition rates. A typical set of stress scans did not reproduce AP crashes. AP error file doesn't reveal any pattern either. Site currently has old style (Reflex-400A) boards. These boards were swapped for the newer Reflex-400B, but there was no change in symptoms. SPR MRIhc06568
Error message
2228010 AP Error 2228016 Internal AP software error See Long Description Please consult file /usr/g/service/log/aperror.log for more information.
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required