GE Healthcare · MRSigna22334042228016

Intermittent Image Reconstruction Failures.

Symptom

Intermittent Image Reconstruction Failures The system intermittently fails to reconstruct images. The failure occurs after the acquisition has completed.

Error message

2233404 VSB Transfer not completed before next transfer scheduled. Wed Jul 15 07:25:04 1998 Error: 2233404 Host: lx-mr Process: NSP File: TPS:cap: c_int58 Line: 5 VSB Transfer not completed before next transfer scheduled. Unsuccessful VSB address: 0x27f8d280.

2228016 Internal AP software error See Long Description Please consult file /usr/g/service/log/aperror.log for more information.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR