Array Processor Errors Due to Master AP Board.
Symptom
Array processor AP errors during scanning. Scans stop (abort) with an AP error logged. It appears that AP crashes are not related to heavy duty recon jobs or high data acquisition rates. A typical set of stress scans did not reproduce AP crashes. AP error file doesn't reveal any pattern either. Site currently has old style (Reflex-400A) boards. These boards were swapped for the newer Reflex-400B, but there was no change in symptoms. SPR MRIhc06568
Error message
2228010 AP Error 2228016 Internal AP software error See Long Description Please consult file /usr/g/service/log/aperror.log for more information.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion MERCURY 910-07072 MINERVA II MASTER 2325947-12 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.