Intermittent Image Reconstruction Failures Due to AP Board.
Symptom
Some or All Images in the Last Scan Did Not Reconstruct Some or All Images in the Last Scan Did Not Reconstruct.
Error message
2228010 AP Error 2228016 Internal AP software error See Long Description Please consult file /usr/g/service/log/aperror.log for more information.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2325947-12 AP Reflex Minerva II 400B Master Board - M3000RJ For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering.
- Area
- MR
Full cause & step-by-step resolution