GE Healthcare · MRExcite224939111267789831374722493518275114048
Will Not Scan with Semaphore Errors Due to APS Board.
Symptom
Will Not Scan with Semaphore Errors Fails during scanning with errors 2249351, 2249391 , 2249392.
Error message
2249391 semTake failed for semaphore ID 0x3FFFDF0 1126778983 2 1 Thu Sep 15 10:09:43 2005 2249391 4 MR3T diagCoord sysIntegrity.c 1010 Task ID: 0x3fffbb0 Time: 13747 Object: APS semTake failed for semaphore ID 0x3FFFDF0. Errno = 0x3D0004. EN 218 SR 218 2249351 Exception: exception type Data Access (DSI). task number xxxxxx; task name xxxx, exception vector xxxx, pc xxxx. Host: b4 Proc: diagCoord Error:2249351 File: excLog.c Line: 670 Task ID: 0xdd9ef0 Time: 82751 Object: SCP Exception: exception type Data Access (DSI). task number 0x6E2B90; task name aceTimer, exception vector 0x300, pc 0x9E0098. Additional information in /usr/g/service/log/vxExLg-14048 2249392 semGive failed for semaphore ID 0x3FFFDF0 Thu Sep 15 10:09:43 2005 2249392 4 MR3T diagCoord sysIntegrity.c 1027 Task ID: 0x3fffbb0 Time: 13747 Object: APS semGive failed for semaphore ID 0x3FFFDF0. Errno = 0x160068. 2252108 Motorola BLD test detected a error APS RAM see long Sat Apr 3 17:17:06 2010 Error: 2252108 Host: SJHMR01 Process: diagCoord File: diags_monitor.c Line: 298 Diagnostic: MOT Test Iteration: 1 of 1 Action: 1 Task ID: 0x0 Time: 1270333026 Object: Host Diag Executive Motorola BLD test detected a error: APS-
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2294300-4 APS BOARD WITH IT DAUGHTER BD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.