GE Healthcare · MRSignaExciteDiscovery100037453372218071

Exam is not in the Browser Customer calling in indicating that the exam was never in the Browser.

Corrupt Image Database.

Symptom

Exam is not in the Browser Customer calling in indicating that the exam was never in the Browser. As it turns out...it was removed with several other studies on the system.

Error message

9017 system error unable to remove /export/home1/sdc_image_pool2/reserved/.IMSseXBAa000S5 I/O error Sat Apr 3 09:42:06 2010 Error: 9017 Host: PHELPS_MRI Process: imserver File: diskmisc.c Line: 260 system error : unable to remove /export/home1/sdc_image_pool2/reserved/.IMSseXBAa000S5 I/O error

100037 Can't create a BImage for /export/home1/sdc_image_pool2/mr_scan/XXXXX.X.X.X Sat Apr 3 09:42:06 2010 Error: 100037 Host: PHELPS_MRI Process: autoview File: avManagedList.cc Line: 754 Can't create a BImage for /export/home1/sdc_image_pool2/mr_scan/45337.6.1.2

2218071 Cannot rename file /export/home1/sdc_image_pool2/reserved/IMSseXBAa000S5/XXXXX.X.X.X errno = 5 Sat Apr 3 09:42:06 2010 Error: 2218071 Host: PHELPS_MRI Process: TIR File: ImageMisc.cc Line: 521 Cannot rename file /export/home1/sdc_image_pool2/reserved/IMSseXBAa000S5/45337.6.1.2 errno = 5

Applies to

Models
Signa, Excite, Discovery
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR