Unable to scan with the MRID OpenSpeed Head coil.

Symptom

Unable to scan with the MRID OpenSpeed Head coil. All of the other coils work fine. See long description. The system is a OpenSpeed with Excite running HFO4 M4 software. New Installation, unable to scan with the MRID OpenSpeed head coil. The other coils they have tried appear to work OK. The system gives Multicoil TR Driver open faults on MCD1, and also TR Driver faults Head path open. The coil config for the Head coil looks good. {sdc@MRIPLAZA}[1] getver Hostname : MRIPLAZA Build number for cclass is OPENMR4.292a_M4_0433.f Build number for driverSupport is OPENMR4.292a_M4_0433.f Build number for hfopsd is OPENMR4.292a_M4_0433.f Build number for insite is OPENMR4.292a_M4_0433.f Build number for install is OPENMR4.292a_M4_0433.f Build number for MrpApps is OPENMR4.292a_M4_0433.f Build number for mx_apps is OPENMR4.292a_M4_0433.f Build number for mx_plat is OPENMR4.292a_M4_0433.f Build number for mx_terra is OPENMR4.292a_M4_0433.f Build number for OS is GEMS_Linux_1.8.9 Build number for patch_OPENMR4_M4_01 is OPENMR4.292a_M4_0435.a Build number for patch_OPENMR4_M4_02 is OPENMR4.292a_M4_0433.f.PA Build number for patch_OPENMR4_M4_03 is OPENMR4.292a_M4_0433.f.PA Build number for patch_OPENMR4_M4_04 is OPENMR4.292a_M4_0433.f.PA Build number for patch_OPENMR4_M4_05 is OPENMR4.292a_M4_0433.f.PC Build number for patch_OPENMR4_M4_06 is OPENMR4.292a_M4_0433.f.PA Build number for patch_OPENMR4_M4_07 is OPENMR4.292a_M4_0433.f.PA Build number for patch_OPENMR4_M4_08 is OPENMR4.292a_M4_0433.f.PA Build number for patch_OPENMR4_M4_09 is OPENMR4.292a_M4_0433.f.PB Build number for pc is OPENMR4.292a_M4_0433.f Build number for PostSdC is OPENMR4.292a_M4_0433.f Build number for TERRA is IOS_M4_20031119_R292a_M4_Patch2 MR Software release: OPENMR4.292a_M4_0433f

Error message

2249149 T/R driver fault, head path open. 2249149 Host: LDEX Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1286 T/R driver fault, head path open. Possibly a disconnected cable between the driver board and the head T/R network, or an open diode in the head T/R switch.

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-se

Applies to

Models
Excite
Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR