PDU Emergency Stop Circuit Trips When Gradient Control Power Circuit Breaker is Turned On at PDU due to Incoming 480 Vol…

PDU Emergency Stop Circuit Trips When Gradient Control Power Circuit Breaker is Turned On at PDU due to Incoming 480 Volt Power to PDU Phase Reversed.

Symptom

PDU Emergency Stop Circuit Trips When Gradient Control Power Circuit Breaker is Turned On at PDU See Long Description The Teal PDU's emergency stop circuit trips when the Gradient Control power circuit breaker is turned on at the PDU, and it can't be reset as long as the Gradient Control CB is engaged. This happens even if the external emergency stop circuit is bypassed by placing the jumper plug on the PDU Control Board's J2 connection. This is an HFO3 to HFO5 upgrade. The computer was not yet booted so no error messages were logged.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Area
MR

Full cause & step-by-step resolution