GE Healthcare · MRExciteSigna2247383685114
Error 2247383 The signal's channel-to-channel variance exceeded the specification for channel Due to UTNS3. 16 Channel S…
Error 2247383 The signal's channel-to-channel variance exceeded the specification for channel Due to UTNS3. 16 Channel Switch Diagnostics Fail.
Symptom
16 Channel Switch Diagnostics Fail
Error message
2247383 The signal's channel-to-channel variance exceeded the specification for channel: Test 16 Channel Switch Output Osc See Long Description Error: 2247383 Host: b5 Process: diagCoord File: rrfRxFruIso.c Line: 2796 Diagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x7f53e50 Time: 685114 Object: APS Test Task Function: Software/Firmware/Config: N/A Test: '16 Ch Switch Output Osc. OB(O)' diagnostic error. The signal's channel-to-channel variance exceeded the specification for channel: 7, 8.-> Extended Error: Review the data in /usr/g/service/datamine. The summary file contains the results used for logging this error. If the file exists, the raw and processed data is available in a corresponding data file whose name contains the test iteration tag.
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2366218-2 UTNS3 See Long Description 2366218-2 UTNS3 EMC ED2 Compliant .
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.