CWA Unable to Scan with Power Monitor Faults due to Split Head Coil Halves Were Not Completely Closed Together.

Symptom

Unable to scan with Power Monitor Faults

Error message

2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 NSP SCP:UPM 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. Peak value detected = 429 W. Max Peak Limit = 426 W- Possible cause(s): ASC RF Detector, ASC RF Processor, RF Amp, Software .

2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.

2254616 RF Amplifier Fault. 2254616 Host: WHITEMRI Process: NSP File: SCP:Srfd3RfNbCanHWO Line: 2289 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Spin Echo Coil: HEAD Scan: 

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.