2252302 Multicoil Bias Fault Due to Coil.

Symptom

Intermittently cannot scan with Error 2252302 Multicoil Bias Fault - Transmit Fault. Try re-seating all coil connectors. See Long Description Intermittently cannot scan with Error 2252302 "Multicoil Bias Fault - Transmit Fault. Try re-seating all coil connectors." This may be happening on multiple coils and ports, including ports not in use or ports nonexistent on the system. They may immediately retry the scan and it may work. But the problem returns later. See example, below. iTrak # 13283625. Error: 2252302 Host: OUMCGEMR1ET Process: NSP File: SCP:DCB Line: 891 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Gradient Echo Coil: HNS Head Scan: 2667/2 Multicoil Bias Fault - Transmit Fault. Try re-seating all coil connectors. MCD2 Channel 15 of 16, counting from 1 Value: -2 mA, Limit: 0 mA Mask of failed channels: 0x4000 MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C->

Error message

2252302 Multicoil Bias Fault - Transmit Fault. 2252302 Multi-coil bias fault - transmit fault. MCD1 channel 3 of 16, counting from 1 Value: -72mA. Limit 0mA Mask of Failed channels: 0x9c MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
SecondOpinion 1.5T 8 Channel TR Knee Array 5114356-2 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.