GE Healthcare · MRExcite2254212
Prescan Fails Due To SRFD2 RF Amplifier Peak Power Fault.
Symptom
Auto and Manual Prescan Fails
Error message
2254212 SRFD2 RF Amplifier Peak Power Fault. Peak Power from RF Amplifier Exceeds Specification Fri Apr 30 14:23:00 2004 Error: 2254212 Host: lx-mr1 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 SRFD2 RF Amplifier Peak Power Fault. Peak power from RF Amplifier exceeds specification(72.5dBm). Possible causes: the PSD exceeds system limits, RF Gain miscalibration, defective Exciter, or defective RF Amplifier.->
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR