Pular para o conteúdo principal
GE Healthcare · MRSignaExcite2224733222533722253342225935

Alignment Lights, Patient Blower and Bore Light Do Not Function Due To SSM.

Symptom

Alignment Lights, Patient Blower and Bore Light Do Not Function Alignment Lights, Patient Blower and Bore Light Do Not Function. PAC error 2225935 in error log also.

Error message

2224733 - The SRI is not responding. Please reset the TPS chassis and try again.

2225337 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Error: 2225337 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri watchdog fail Line: 13 SPI-SRI Communication Error. SRI failed to respond to a watchdog packet from SPI. Watchdog packets waiting for a response: 0x0001

2225334 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Error: 2225334 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri resp timeout Line: 12 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Data packets waiting for a response: 0x4000

2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. 2225935 The PAC does not respond. Check PAC power and fiber optic connections, run PAC diags. This error condition adversely affects gated scans only. 2225935 The PAC does not respond. Check PAC power and fiberoptic connections, run PAC diags. This error condition adversely affects gated scans only.

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues