GE Healthcare · MRExciteSigna224505359528
Will Not Scan with Gradient Amplifier % Axis Current Distortion Fault due to BRM-D
Symptom
System Will Not Scan With Gradient Amplifier Current Distortion Faults.
Error message
2245053 GRADIENT DRIVER TESTS - The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault. Ready Static Mode-> 2245053 lx-mr diagCoord grad_lib.c 921 Diagnostic: GDT Static Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c6c8 Time: 59528 Object: SCP Test Task GRADIENT DRIVER TESTS - The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault. Ready Static Mode->
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion BRM-D 2144093-5 See long description for SMILe on returning BRM's. See comments in resolution concept for second opinion requirements before ordering this part SMILe 2/7/2006 Subject: Failed and unused BRM coils must be returned promptly for evaluation and repair. Effectivity: All Signa 1.5T Horizon, LX, EXCITE and HD MR systems. Problem: Unused BRM's are remaining on site too long which severely limits the availability of FRU BRM coils. The repair facility is able to perform quick evaluations/repair of coils but BRM's must be returned promptly. Resolution: BRM coils should be ordered with no more than a 3-day lead time . Every effort is being made to always have FRU BRM's available but your help is needed. Within 10 days of shipment of the FRU BRM either it or the failed unit must be returned. Any exceptions to this time limit must be cleared through Ramamohan Pai, Global Modality Leader-MR & XR. Email : [email protected] Tel: 1-262-544-3240, Dial Com : 320-3240
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.