Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova281001016100262800077

Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.

Symptom

Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76

Error message

INX1PRO-IQ+: Error code 2810010 Update of console status has failed INX1PRO-IQ+: 2810010 Update of console status has failed IH XRSubSysMgrDL Failed to update status on console error @ StatusUDPListener.cxx 406. See long description for further details on full pattern. INX1PRO-IQ+:only date April/02/2010 Inhibit error 2810010. System impact : X-ray not available anymore. Sometimes, system does not recover after reset at VCIM. Customer needs to power off, then power on the complete system. Full pattern for this problem is the sequence of errors below: - 1610026 Front: Frontal Capture Board Error DO XRImProc DriverFcib::_checkErrIT() -> Fiber channel receiver SEQUENCE-B error - 2810010 Update of console status has failed IH XRSubSysMgrDL Failed to update status on console error @ StatusUDPListener.cxx 406 - 2800077 ALARM (!! , 2.1) - Communication lost between DL and RTAC WR XRSubSysMgrDL Ethernet link to RTAC has been lost Detailed pattern seen through Proactive Digital Service (PDS):

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
$ INX1PRO or IQ+,Harmony_M3 INX1PRO or IQ+, Harmony_M3 Product is called INX1PRO or IQ+: Harmony_M3 sw release applicable to Harmony systems.
Component
INX1PRO-IQ+: RTAC 8 bootable chassis RTAC 8 bootable chassis. See long description for FRU reference number. RTAC 8 bootable chassis FRU reference 5268791. Always use the most appropriate documentation for your system. The following FRU reference was taken from Direction 5311070-2-2EN, Rev# 4 and is provided for your reference from the Innova 2100-IQ, 3100-IQ, 4100-IQ Harmony DVD documentation. To view the most recent appropriate and detailed documentation, go to MY WORKSHOP
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required