Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova1050015

1050015 C-arc AXIS POT ENCODER MISMATCH

Symptom

C-arc not moving Customer said the gantry motion did not work anymore. Then, he told that initial problem was located on the C-arc motion. He was unable to drive C-arc (motion starts then automatically stops). If L-arm at 0°, only C-arc motion is impacted whatever machine / anatomical mode selected on Smart handle / smart Box. if L-arm is moved to a position different from 0°, C-arc AND pivot motion will be impacted in ANATOMICAL mode. Customer said the gantry motion did not work anymore. Then, he told that initial problem was located on the C-arc motion. He was unable to drive C-arc (motion starts then automatically stops). If L-arm at 0°, only C-arc motion is impacted whatever machine / anatomical mode selected on Smart handle / smart Box. if L-arm is moved to a position different from 0°, C-arc AND Pivot motion will be impacted in ANATOMICAL mode.

Error message

1050015 C-arc AXIS POT ENCODER MISMATCH DO GantryAxis Intermittent Motion stop C-arc AXIS POT ENCODER MISMATCH 1 0 - potentiometerEncoderMismatch Z:/heroic/motion control task.c 4560 1050015 C-arc AXIS POT ENCODER MISMATCH DO GantryAxis Intermittent Motion stop C-arc AXIS POT ENCODER MISMATCH 1 0 - potentiometerEncoderMismatch Z:/heroic/motion control task.c 4560

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
$ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
Component
gantry arc pot gantry arc pot gantry arc pot
Area
X-ray Vascular

Full cause & step-by-step resolution