Unable to Select New Patient, Patient Register button for New Pt Is Grayed Out, due to corrupt config files.
Symptom
Unable to Select New Patient, Patient Register button for New Pt Is Grayed Out, Not Available Cannot Select New Patient, "Patient Register" button for "New Pt" Is Grayed Out, Not Available Normal appearance of New Patient button would look as follows:
Error message
2248681 MDS link peripheral file gp.t31 failed to load. Tue Mar 2 11:43:04 2010 Error: 2248681 Host: MR_LMC Process: NSP File: SCP:MdsAutoConfig Line: 774 MDS link peripheral file gp.t31 failed to load. Please fix the file and perform a TPS Reset. 2223836 Download APS failed, ldLoadAt() system call failed Tue Mar 2 11:42:59 2010 Error: 2223836 Host: MR_LMC Process: NSP File: TardisReset.cc Line: 1189 Download APS failed, ldLoadAt() system call failed. 2248518 An error occurred trying to read the file gp.t31.Please fix and perform a TPS Reset Tue Mar 2 11:43:04 2010 Error: 2248518 Host: MR_LMC Process: NSP File: SCP:PeripheralImage Line: 185 An error occurred trying to read the file gp.t31.Please fix and perform a TPS Reset.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.