GE Healthcare · MRExciteSigna224914722246662224583
Low Voltage on the High Voltage SSM Supply.
Symptom
Getting SSM Error When Auto Prescanning
Error message
2249147 High Voltage Power Supply Low. Also may be a shorted load on dynamic disable or direct drive circuits. Tue Mar 30 13:50:47 2004 Error: 2249147 Host: CCRC Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1255 SSM fault.^A High voltage power supply low. Also may be a shorted load on dynamic disable or direct drive circuits. 2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure. 2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.