MRIQ Shading in Images Due to the Receive Coil in use.
Symptom
MRIQ Shading in Images See long description Image Parameters Exam Description L-SPINE Series Description T1 SAG Anatomic Reference SN Scan Protocol Name Patient Weight (kg , Lb) 86 , 190 Image Type GE Advantage Windows Image Format Rectangular Image Matrix (pixels) 512 Image Suite ID #GEMS Study number 3128 Series number 3 Image number 6 Series Date & Time Nov 19, 1998 9:41:41 AM Image Date & Time Nov 19, 1998 9:41:41 AM Contrast information NO C Display FOV (mm) 300 Slice Thickness (mm) 4.0 Magnet Strength (Gauss) 15000 Number of excitations 2.0 Number of echoes 1 Flip Angle for GRASS(°) 90 TR (ms) 500 TI (ms) 0 TE (ms) 13 TE2 (ms) 0 Center Freq.(0.1Hz) 638915050 Auto Center Freq(0.1Hz) 638915050 Auto Trans.Gain(0.1dB) 170 Prescan R1 Analog 11 Prescan R2 Digital 15 SAT fat/water/none 0 Variable Bandwidth(Hz) 15.625 Nb of slices in this group 12 Phase contr. flow axis 0 Phase contr. velocity encoding 0 Prescan Flag Auto Prescan Succeeded Pulse Seq. Name fse-xl PSD Name FSE-XL Coil Name CTLBOT Slice Location (mm) -23.3 Scan Spacing (mm) 1.0 Recon. x (mm) -23.3 Recon. y (mm) -18.0 MRIQ Customer Image Library
Error message
2224709 Low signal detected. Please check the signal level for receiver x or receiver x+1 while using Manual Prescan. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver 1 (or receiver 2 while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver (X + 1) while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 5146634-2 1.5T HD 8CH HiRes Brain Coil with Baseplate and Mirrors 5146634-2 1.5T HD 8CH HiRes Brain Coil with Baseplate and Mirrors.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.