GE Healthcare · MRExcite2249148
Cannot Scan With intermittent TR and Dynamic Disable Faults.
Symptom
Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.
Error message
2249148 Dynamic disable and T/R driver board error. A spurious T/R or inductive drive fault occurred See Long Description 2249148 ERROR: EM_HC_TR_MON_REG_2_FLT Dynamic disable and T/R driver board error. A spurious T/R or inductive drive fault occurred. It cleared itself before the specific problem could be read. Possible causes: noise onthe dynamic disable board or a bad board.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.