GE Healthcare · MRSignaExcite2244653

2244653 The GSC indicated a CPU reset during normal operation Due to the Gradient Switch Module.MRIQ Gradient Switch Rel…

2244653 The GSC indicated a CPU reset during normal operation Due to the Gradient Switch Module.MRIQ Gradient Switch Related Errors Pause Scans and Resultant Images Contain a Bright Zipper or Band Art…

Symptom

MRIQ Gradient Switch Related Errors Pause Scans and Resultant Images Contain a Bright Zipper or Band Artifact Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. See the Attachment for copies of the errors and an example image. Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. See the Attachment for copies of the errors and an example image. This system is a 1.5T Excite II TwinSpeed running 11.0 M4 software with Service Pack 1 loaded. Scans suddenly pause then the system lights the Start Scan button for the customer to restart the scan. If this is done the images show a big zipper-like artifact that looks like a partially collapsed gradient. This is seen on many different scans not unique to one PSD. The problem may occur several times a day, and it may skip some days altogether, then reappear again another day. There are several error messages logged which seem to point to the GSC (Gradient Switch Controller). Below are copied the error messages logged at the time of one of these scan pauses and a typical image. The fix was to replace the "TWIN GRAD SWITCH BOX" part number 2253466 in the TAC cabinet. This module contains the GSC. Reference SPR # MRIhc16978. Mon Sep 25 10:12:27 2006 2244648 WHMR1 NSP SCP:GP 668 The GSC reported a problem with the Whole switch on the Y axis. If the problem persists, please call a service engineer. Mon Sep 25 10:12:28 2006 2244655 WHMR1 NSP SCP:GP 1307 The GP received an ALARM message from the GSC. Subsequent error messages should reveal the cause of the alarm. Mon Sep 25 10:12:35 2006 2244598 WHMR1 NSP SCP:GP 1069 The gradient system on the Y axis recovered successfully. There are no faults at the present time. Mon Sep 25 10:12:35 2006 2244598 WHMR1 NSP SCP:GP 1069 The gradient system on the Z axis recovered successfully. There are no faults at the present time. Mon Sep 25 10:12:35 2006 2244598 2 WHMR1 NSP SCP:GP 1069 The gradient system on the X axis recovered successfully. There are no faults at the present time. Author Paul Lyman SME John Schumacher

Error message

2244653 The GSC indicated a CPU reset during normal operation. TPS reset may be necessary. See Long Description Error: 2244653 Host: GE3TMR Process: NSP File: SCP:Grad Line: 683 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FRFSE-XL Coil: HDCTL 456 Scan: 190/3 The GSC indicated a CPU reset during normal operation. TPS reset may be necessary. If this error is persistent, check the GSC for loose or dirty internal power connections. Check communication connections.

Applies to

Models
Signa, Excite
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues