GE Healthcare · MRExciteSigna22446482244655
2244648 The GSC reported a problem with the Whole switch on the Y axis due to Gradient Switch Module.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244648 The GSC reported a problem with the Whole switch on the Y axis. If the problem persists, please call a service engineer. 2244648 WHMR1 NSP SCP:GP 668 The GSC reported a problem with the Whole switch on the Y axis. If the problem persists, please call a service engineer. 2244655 Gradient Switch Fault Detected. Scan will be paused 2244655 Host: TwinExciteLXMR Process: NSP File: SCP:Grad Line: 704 Gradient Switch Fault Detected. Scan will be paused.-> Extended Error: Please restart the scan (continuing the scan might cause image quality issues). If the fault persists,replace the Gradient Switch
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.