Unable to Scan with Gradient Errors Due to 5 Volt MGD Power Supply.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2244577 The GP reported a % Axis Current Distortion Fault 2244577 "The GP reported a Z Axis Current Distortion Fault. If no other faults occurred, check LCoil Calibration. Else, run Gradient Driver Tests to isolate the FRU. If Tests pass, see Troubleshooting Manual." 2248267 The RHO_DDS freq was set to 0 Hz. This value is too small. 2248267 11X, E2, G3,OPENMR4,5 error code is changing to 2248293 Post 11.1, E2M4a, G3M4b, OPENMR4, 5 sw 1121284117 6 1 Wed Jul 13 19:48:37 2005 2248267 4 lx-mr NSP SCP:cex_set_center_freq 1207 The RHO_DDS freq was set to 0 Hz. This value is too small. RHO_DDS will be set back to its default value. Probably a software error. 2254012 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Rx Signal Detect signal is in an Invalid State. Tue May 14 19:20:37 2002 2254012 1 bay12b_oc0 NSP SCP:RRF-DIF 630 RRF-DIF Fiber Optic Link Error on the Master RRF-DIF board. Rx Signal Detect signal is in an Invalid State.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2292267 MGD RF Chassis 5 Volt Power Supply in front of MGD ----RF chassis -------
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.