GE Healthcare · MRExcite224826613443420162248288
MRIQ Black Bands in Images Due to IRF I/O Board.
Symptom
MRIQ: Black Bands through anatomy See the attached example image.
Error message
2248266 File APS DRF APS received DRF utopia rx bus checksum error. Possible reason: IRF/DRF hardware error. Please perform a TPS Reset. DRF ID:1 FE ID:0 Int Status: 1344342016 Counter: 1-> 2248288 post 12M5a, 14M5 sw is the same error as 2248266 on 12M4, 12M5, 14M4 sw.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution