Table up motor Runs Continuously Due to Broken Dock Pedal Spring.Table up motor Runs Continuously after it's all the way…

Table up motor Runs Continuously Due to Broken Dock Pedal Spring.Table up motor Runs Continuously after it's all the way up.Replace Dock Pedal Spring.Excite HD.12.0

Symptom

Table up motor Runs Continuously after it's all the way up

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
46-271152P1 Table Dock Pedal Spring 2 10.42" AUSTENTIC STN. STEEL WIRE