System is Logging 2277671 ICN sensor detects value out of range, but System is Running Fine.
Symptom
System is Logging 2277671 ICN sensor detects value out of range, but System is Running Fine See Long Description The system is logging the following error once every hour, for each of the system's four ICNs. Other than logging the error, there doesn't seem to be any other impact on the system. The system is running fine. Should we troubleshoot this error? SPR # MRIhc45772 Error: 2277671 Host: Signa-3 Process: diagCoord File: /export/home/signa/vre/diags/sen Line: 614 Exception Class: Notification Function: Software/Firmware/Config: N/A ICN sensor detects value out of range. ICN: [icn2 MAC=00:30:48:58:93:26 SERIAL=1495807] Sensor: 1.8V Higher Limit: 0 Current Value: 1.836->
Error message
2277671 ICN sensor detects value out of range See Long Description Error: 2277671 Host: Signa-3 Process: diagCoord File: /export/home/signa/vre/diags/sen Line: 614 Exception Class: Notification Function: Software/Firmware/Config: N/A ICN sensor detects value out of range. ICN: [icn2 MAC=00:30:48:58:93:26 SERIAL=1495807] Sensor: 1.8V Higher Limit: 0 Current Value: 1.836-> Extended Error: 2277671 ERROR: EM_DIAG_ICN_SENSOR_OUT_OF_RANGE ICN sensor detects value out of range. ICN: %s Sensor: %s %s Limit: %s Current Value: %s Extended Error: Check detected hardware of ICN Possible cause(s): Img Compute Node
Applies to
- Models
- Excite, Signa
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.