Pular para o conteúdo principal
GE Healthcare · MRSignaOvationExcite223340422185372233425

CWA Scan Fails to Complete with The 3D Pass Done Buffer Queue has overflowed.

Symptom

Scan Fails to Complete with The 3D Pass Done Buffer Queue has overflowed. May need a longer TR time, or more time between echoes Scan Fails to Complete with The 3D Pass Done Buffer Queue has overflowed. May need a longer TR time, or more time between echoes.

Error message

2233404 VSB Transfer not completed before next transfer scheduled. Wed Jul 15 07:25:04 1998 Error: 2233404 Host: lx-mr Process: NSP File: TPS:cap: c_int58 Line: 5 VSB Transfer not completed before next transfer scheduled. Unsuccessful VSB address: 0x27f8d280.

2218537 Scan Stopped, by Cerd Error Server Request Fri Jan 18 16:01:10 2002 Error: 2218537 Host: lx-mr Process: scn File: MsaScanMgr.m Line: 2487 Scan stopped, by Cerd Error Server request.

2233425 The 3D Pass Done Buffer Queue has overflowed. May need a longer TR time, or more time between echoes Thu Apr 2 15:58:18 2009 Error: 2233425 Host: lx-mr Process: NSP File: TPS:cap: c_int24 Line: 1022 The 3D Pass Done Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan.

Applies to

Models
Signa, Ovation, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.