2254201 SRFD2 RF Amplifier General Overheating Fault caused by SRFD2 RF Amplifier

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2254201 SRFD2 RF Amplifier General Overheating Fault 2254201 Host: GEMR1 Process: NSP File: SCP:../rf nb contro Line: 1491 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Unknown Coil: Unknown Scan: -1/-1 SRFD2 RF Amplifier General Overheating Fault. RF Amplifier has exceeded maximum internal temperature. Possible causes: restricted air flow, defective fans, or defective SRFD2 combiner

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573