Unable to Scan using the USAI Large Knee Coil with Error Code 2252301 due to the Coil's Cable Was Damaged Creating an Op…
Unable to Scan using the USAI Large Knee Coil with Error Code 2252301 due to the Coil's Cable Was Damaged Creating an Open Circuit.
Symptom
Unable to Scan using the USAI Large Knee Coil Legend 5000 Phased Array Knee and Foot Coil with 2252301Multicoil Bias Fault Detected Open Circuit See Attachment for all details.
Error message
2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-seat all coil connectors. The mask of all failed channels indicates additional channels with the same fault - convert the hex number to a bitmap to decipher the mask, with channel 1 indicated by the least significant bit and channel 16 indicated by the most significant bit. For example, a mask of 0x3c converts to 0011 1100, meaning channels 3, 4, 5, and 6 all experienced the same failure. The key at the bottom of the error message shows the MC driver bias channels provided by the Driver Module used with each mult
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR5-HFO5 Open Speed with Excite Product
- Component
- 2264742-14 0.7T EXTREMITY COIL CABLE--LARGE
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Prescan Fails with T/R driver fault, head path open Errors Due to Intergrated Head Assembly (formerly Head TR Switch)
- T/R Driver Fault is Due to Broken Pin on 5 inch Round General Purpose Coil.
- Multicoil Bias Fault errors on multiple coils.
- Auto Prescan Failures and RF amp setup incomplete Errors Due to Defective RF amp.
- The system is unable to scan with any surface coil, get dynamic disable faults.