Pular para o conteúdo principal
GE Healthcare · MRExcite2249149

T/R Driver Fault is Due to Broken Pin on 5 inch Round General Purpose Coil.

Symptom

Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.

Error message

2249149 T/R driver fault, head path open. 2249149 Host: LDEX Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1286 T/R driver fault, head path open. Possibly a disconnected cable between the driver board and the head T/R network, or an open diode in the head T/R switch.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2127316 5 inch Round General Purpose Coil
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.