GE Healthcare · MRExcite22238492252700
TPS download problem Due to the network switch in the System Cabinet.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2223849 AGP EPROM power-up boot failed See Long Description Error: 2223849 Host: rmns Process: NSP File: TardisReset.cc Line: 1781 AGP EPROM power-up boot failed. Communication Timed-out from AGP board. Verify that MCP820 boards are installed and seated properly for APS and AGP. Extended Error : MCP820 boards can be identified by the word MCP820 printed on the board extraction lever. 2223849 SCP EPROM power-up boot failed (See attachment for extended error message) 2223849 rmny NSP TardisReset.cc 1841 Exception Class: Unknown Function: Host: NSP SCP EPROM power-up boot failed. Communication Timed-out from SCP board. Verify that MCP820 boards are installed and seated properly for APS and AGP.-> 2252700 Software Error: A call to a VxWorks function returned an error Thu Sep 15 10:09:48 2005 2252700 4 MR3T NSP AGP:spu_if.c 162 Software Error: A call to a VxWorks function returned an error. VxWorks Function: semTake Error: S_objLib_OBJ_TI
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.