GE Healthcare · MRExciteSigna22488502248822
Auto Prescan failed, RF Amplifier is not ready Due To Defective RF Amplifier.
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2248850 RF Amplifier fault 71 Mon Nov 27 10:10:04 2006 Error: 2248850 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 RF Amplifier fault 71. 2248822 Analogic RF Amplifier 1 High Voltage Low Fault Mon Nov 27 10:03:56 2006 Error: 2248822 Host: VDMRI Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 Analogic RF Amplifier 1 High Voltage Low Fault. Analogic Amplifier 1 power supply voltage dropped below 120V DC.(150V nominal).->
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.