GE Healthcare · X-ray VascularInnova1000025
Positioner and Jedi not communicating to RTAC.
Symptom
System Boot Failed: Boot Timed out: Innova boot issue System Boot Failed: Boot Timed out: The Innova failed to boot properly.
Error message
Error Code = 1000025, error Text = ARCNET Communication Failure, debug Text = Positioner link disconnected, recovery Class = IH, major Function /= XRImChainPos Error Code = 1000025, error Text = ARCNET Communication Failure, debug Text = Positioner link disconnected, recovery Class = IH, major Function /= XRImChainPos. System impact: X-ray and Gantry motion unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service".
Applies to
- Models
- Innova
- Product
- X Innova 4100 X Do not use---------4100 systems 4100 systems
- Software
- $ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.