2244649 The GSC reported a problem with the Zoom switch Due to the Gradient Twin Switch Module.Scanner Pausing Between S…

2244649 The GSC reported a problem with the Zoom switch Due to the Gradient Twin Switch Module.Scanner Pausing Between Series when Switching from Whole to Zoom Mode.Replace the Twin Gradient Switch Mo…

Symptom

Scanner Pausing Between Series when Switching from Whole to Zoom Mode

Error message

2244649 The GSC reported a problem with the Zoom switch on the Y axis

2244655 Gradient Switch Fault Detected. Scan will be paused 2244655 Host: TwinExciteLXMR Process: NSP File: SCP:Grad Line: 704 Gradient Switch Fault Detected. Scan will be paused.-> Extended Error: Please restart the scan (continuing the scan might cause image quality issues). If the fault persists,replace the Gradient Switch

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2253466 TWIN GRADIENT SWITCH Module Box Same part number for 1.5T and 3.0T. Same part number for 1.5T and 3.0T.