Pular para o conteúdo principal
GE Healthcare · MRExcite2224580

The signal is too small due to damaged quad knee coil cables.

Symptom

Auto Prescan Fails With Knee Coil

Error message

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2383613-3 Medical Advances Quad Knee Coil Cable Assembly MED ADV QUAD KNEE CABLE.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues