Pular para o conteúdo principal
GE Healthcare · MRSignaExcite224486922448712244872

Y Gradient OverTemperature and OverCurrent and OverVoltage Faults due to Control Board.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244869 The GP reported multiple % Axis OverTemperature Faults. Tue Dec 29 07:39:40 2009 Error: 2244869Host: lx-mr Process: NSPFile: SCP:Grad Line: 2681Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1The GP reported multiple Y Axis OverTemperature Faults.-

2244871 The GP reported multiple Gradient Amplifier % Axis OverVoltage Faults. See Long Description Extended Error : If an Overvoltage fault is not also reported on the SGA Power Supply, replace the Gradient Amplifier. Else replace the SGA Power Supply. TPS reset is required to recover from this fault.

2244872 The GP reported multiple Gradient Amplifier % Axis OverCurrent Fault. Tue Dec 29 07:39:40 2009 Error: 2244872Host: lx-mr Process: NSPFile: SCP:Grad Line: 2728Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1The GP reported multiple Gradient Amplifier Y Axis OverCurrent Fault.-

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues