GE Healthcare · X-ray VascularInnova1200114

INX100 Unity 2 min Delay in Innova system boot.

Symptom

INX100 Unity 2 min Delay in Innova system boot. Error 1200114 Detector boot warning in log. Delay in system boot on turn on or reset. Error 1200114 Detector boot warning in log 4 time followed in 2 min by a clear. System then boots OK.

Error message

1200114 ERR_DETECTOR_BOOT_WARNING Temperature out of range : x.xx degrees 1200114 ERR_DETECTOR_BOOT_WARNING Temperature out of range : x.xx degrees. The Detector temperature is reported out of Range. Review the temperature of the detector reported in the debug text for further information. Also the detector temperature can be reviewed by performing a telnet to the RTAC and typing dem_status at the prompt. This will give detector power supply status, temperature, and humidity monitoring.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
$ Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9 FMI 12105 Unity M4 SP3.2 DL_Unity_4.18.9, FMI 12105 healthpage tag for this software is UNITY_M4_FMI_4.21 212009270 03/26/2010 13:31 GMT
Area
X-ray Vascular