Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova105017910502601050261105026210501031050042

Innova; Table and Gantry will not move by any control.

Symptom

Innova; Table and Gantry will not move by any control. Table and Gantry cannot Move, TSSC prd display: BD Innova; Table and Gantry will not move by any control.

Error message

error Code = 1050179, error Text = EMERGENCY STOP, debug Text = - EMERGENCY STOP OCCURRED. PRESS RESET UNDER TSSC TO CONTINUE (4AED), recovery Class = WR, major Function /= GantryAxis System impact: Gantry lock-up. Displayed message : "Emergency stop occurred: press reset button to continue".

1050260 FRT LARM AXIS SWITCH ENGAGED -Limit switch 2 problem (prd: BC) 1050260 FRT LARM AXIS SWITCH ENGAGED WR GantryAxis No Motion LARM_Limit_Switch 1 0 - Limit switch 2 problem (prd: BC) Z:/heroic/prd_task.c 638

INX100: 1050261 FRT PIVOT AXIS SWITCH ENGAGED (PRD BC) error code 1050261 FRT PIVOT AXIS SWITCH ENGAGED WR GantryAxis No Motion PIVOT_Limit_Switch 1 0 - Limit switch 2 problem (prd: BC) Z:/heroic/prd_task.c 638 MCB prd code BCH is displayed on the MCB is displayed in the error log. The Pivot MCB has a PRD code of BC displayed on it. INX100: 1050261 FRT PIVOT AXIS SWITCH ENGAGED (PRD BC) error code 1050261 FRT PIVOT AXIS SWITCH ENGAGED WR GantryAxis No Motion PIVOT_Limit_Switch 1 0 - Limit switch 2 problem (prd: BC) Z:/heroic/prd_task.c 638 MCB prd code BCH is displayed on the MCB is displayed in the error log. The Pivot MCB has a PRD code of BC displayed on it. 1050261: Front: FRT 

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues