Pular para o conteúdo principal
GE Healthcare · MRExciteDiscoveryOptima222384920000235222494182000023792000023722223880

CWA Application Software does not complete the boot process with TPS Download failure.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2223849 SCP EPROM power-up boot failed (See attachment for extended error message) 2223849 rmny NSP TardisReset.cc 1841 Exception Class: Unknown Function: Host: NSP SCP EPROM power-up boot failed. Communication Timed-out from SCP board. Verify that MCP820 boards are installed and seated properly for APS and AGP.->

200002352 A HOST_SHUTDOWN was initiated by the death of RIR. Error: 200002352 Host: LATT15 Process: hostMonitor File: ssProcStop.c Line: 812 A HOST_SHUTDOWN was initiated by the death of RIR.

2249418 Network problem detected. Check MGD subnet network configuration. See Long Description Wed Apr 14 09:27:10 2004 2249418 2 CARD diagCoord mgdArp.c 157 Task ID: 0xf99df8 Time: 981 Object: SCP Network problem detected. Check MGD subnet network configuration. etherAddrResolve failed for network interface motfec0, IP 10.47.1.2, numTries 5, delay 100 ticks. status = -1.

200002379 Process RIR, pid 4877 exited with status=255 Error: 200002379 Host: LATT15 Process: hostMonitor File: ssSigHandlers.c Line: 136 Process RIR, pid 4877 exited with status=255

200002372 The process WavRec took too long to start up.

200002372 The process RIR took too long to start up.

2223880 Reset TPS w

Applies to

Models
Excite, Discovery, Optima
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Area
MR

Full cause & step-by-step resolution