GE Healthcare · MRExciteOvationSigna2223883
Application Software does not complete the boot process with TPS Download failure.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2223883 VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs See long description Mon Jan 26 08:47:14 2009 Error: 2223883 Host: OVMRI1 Process: NSP File: VREReset.cc Line: 937 Exception Class: Unknown Function: Recon: Data Acquisition VRE reset failed due to failure to communicate with all ICNs! Check Ethernet and IPMI connectivity to all ICNs.-> Extended Error: The TPS reset failure is caused by the failure to communicate with all configured ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.
Applies to
- Models
- Excite, Ovation, Signa
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Component
- 5130108 Value Image Compute Node Value ICN used on HDe, ProfileHD and OvationHD
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.