GE Healthcare · MR
Magnet Monitor 3 was not placed into Service Mode prior to Magnet Service.hfo_orgi2 alarm I2 - Recon Ruo CBL Open
Symptom
hfo_orgi2 alarm I2 - Recon Ruo CBL Open hfo_orgi2 alarm I2 - Recon Ruo CBL Open, the Recon RUO cable has been disconnected or is damaged in a way that the Magnet Monitor 3 is detecting an open circuit.
Error message
hfo_orgi2 alarm I2 - Recon Ruo CBL Open hfo_orgi2 alarm I2 - Recon Ruo CBL Open, the Recon RUO cable has been disconnected or is damaged in a way that the Magnet Monitor 3 is detecting an open circuit.
Applies to
- Product
- Magnet HFO. Software Version Concept should NOT be selected for Magnet Products Serial # (Txxx) Field strength: 0.7 T He Vessel Pressure >0.9 psig but <6.25 psig Shield Temp for Cryo-cooler coldhead < 40 K Temp for Shield-cooler Coldhead < 35 K Cryocooler 2nd stage Temp > 3.8K but < 4.5K Recondenser Temp > 4.2K but < 4.6K Delta Temp (Coldhead - Recondensor) < 0 but < 0.3K Min. LHe Fill Level 75% of total vessel Min. LHe Ramp Level 100% Helium Volume 340 Liters Burst Disk 2253802-2 Baffle Assy 2226764
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.