XRAY UNAVAILABLE; CLOSE DOOR TO CONTINUE is displayed on the monitor.
Symptom
INX100: XRAY UNAVAILABLE; CLOSE DOOR TO CONTINUE is displayed on the monitor. Error 200040 INX100: XRAY UNAVAILABLE; CLOSE DOOR TO CONTINUE is displayed on the monitor. Error 200040. You can also see the following errors based upon the system adn how it is being reported by teh RTAC. 13204582 INNOVA 4100 - Intermittant Room Door Interlock Exposure Lockout 13206198 Liberty 3100 - Close Room Door To Continue 13206866 Liberty 4100- Close room door to continue message forcing system reset 13207448 Unity 4100 - Close room door to continue message forcing system reset INX100: XRAY UNAVAILABLE; CLOSE DOOR TO CONTINUE is displayed on the monitor. Error 200040. You can also see the following errors based upon the system adn how it is being reported by teh RTAC. 13204582 INNOVA 4100 - Intermittant Room Door Interlock Exposure Lockout 13206198 Liberty 3100 - Close Room Door To Continue 13206866 Liberty 4100- Close room door to continue message forcing system reset 13207448 Unity 4100 - Close room door to continue message forcing system reset Following engineering investigation of this problem, it is found that the issue is due to malfunction in the RTAC software and a fix for this issue is scheduled and will be available in the next installed base software release ( by FW9 2009).
Error message
200040 XRAY UNAVAILABLE; CLOSE DOOR TO CONTINUE Eror in Envent log, 4 ( 4 = 20004 for XRAcqSync) PrepRefused ~ Type_Of_Inhibit = ACQ_INHIBIT ~ Condition_Id_List = (2 ROOM_DOOR) 200040 XRAY UNAVAILABLE; CLOSE DOOR TO CONTINUE Eror in Event log, 4 ( 4 = 20004 for XRAcqSync) PrepRefused ~ Type_Of_Inhibit = ACQ_INHIBIT ~ Condition_Id_List = (2 ROOM_DOOR) The following errors may reported by the RTAC. The Error varies by system type. 13204582 INNOVA 4100 - Intermittant Room Door Interlock Exposure Lockout 13206198 Liberty 3100 - Close Room Door To Continue 13206866 Liberty 4100- Close room door to continue message forcing system reset 13207448 Unity 4100 - Close room door to continue message forcing system reset For Unity SP3.2 and Lower systems: Following engineering investigation of this problem, it is found that the issue is due to malfunction in the RTAC software and a fix for this issue is scheduled and will be available in the next installed base software release ( by FW9 2009).
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- DUAL FOOTSWITCH and cable: Part #: 5137930 Qty: 1 Part Desc: POST EVEREST DUAL FOOTSWITCH Part #: 5137930 Qty: 1 Part Desc: POST EVEREST DUAL FOOTSWITCH
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.