Pular para o conteúdo principal
GE Healthcare · MRExciteSigna678095325169634376

Operating System disk will not load on HP xw9300 Host PC.

Symptom

Operating System disk will not load on HP xw9300 Host PC when trying to install FMI 67809 HDe Ver15.0 M4 Release See Long Description Operating System disk (CTT Linux OS 6.3.8 DVD p/n 5325169) will not load on the xw9300 Host PC when trying to install FMI 67809 "HDe Ver15.0 M4 Release" at step 1.3.2.3. There is no error message displayed, the problem is that the " boot: " prompt never appears. The FE obtained a 2nd copy of the OS DVD and retried with the same results. (The FE was verified to be using the proper OS disk that came with the kit). The FE took the same OS DVD to another HDe system with an HP xw8400 computer and it loaded fine on that system. The FE opened a Trackwise report, # 634376.

Applies to

Models
Excite, Signa
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
5143798-15 HP xw9300 Host PC for HDe See Long Description 5143798-15 HP xw9300 Host PC for HDe In FEMC the part description is simply "MR VALUE 1.5T Host".
Area
MR

Full cause & step-by-step resolution