Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedVCT260107236260107460260107203260107705

BrightSpeed HPower, Velocity or Delight HPower

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260107236 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. See Long Description for full error Mon May 2 09:25:58 2005 Error: 260107236 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 2385 Mon May 2 09:25:58 2005 Host : Table Gantry Processor Ermes # : 260107236 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 2385 Function : AXIAL CONTROL : Axial Control Scan Type : Helical Scan: 19/1/1 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. The Encoder and the Drive MUST be Sync'd/Referenced. (Cur State: 10 Cmd: 21) HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Error: 260107460 The Axial AC Drive reported fault code: F:60 Encoder Loss See LD for more details Fri Jul 21 14:29:10 2006 Error: 260107460 Host: tgp Process: tgp subsys File: AX2fault.c Line: 195 Host : Table Gantry Processor Ermes # : 260107460 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications The Axial AC Drive reported fault code: F:60 Encoder Loss This indicates that the Axial drive can not sense the Axial encoder. 

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5182194-2 SL4 ENCODER ASM, with-LED, with MD gear used for Brightspeed Excel. See Long Description for part details. Part #: 5182194-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SL4 ENCODER ASM, with-LED, with MD gear FRU: Yes Status: Orderable Comments: added per to ECR2050449 by sso#305010201
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues