Pular para o conteúdo principal
GE Healthcare · MRExcite2220048

ASSET Calibration Scan Fail.

Symptom

ASSET Calibration Scan Fail See Long Description The system failed only with this type of scans, everything else worked fine.

Error message

2220048 Asset.xml file might be corrupt or invalid combination of coil, plane and application. Host: MRISB1 Process: scn File: SrxPhaseAsset.m Line: 131 Asset.xml file might be corrupt or invalid combination of coil, plane and application.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR