HDx TPS Reset Failure Due to a Image Compute Node ICN.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2223883 Ermes error 2223883 4 lx-hdmr NSP VREReset.cc 932 Exception Class: Unknown Function: Ermes error, key: 32, error: 4009. Ermes error, key: 2223883, error: 4009.

2277651 ICN Hard Disk Health Diagnostic failed. Mon Feb 23 06:44:11 2009 Error: 2277651 Host: EPMR02 Process: diagCoord File: icnDiskHealth.pl Line: 363 Exception Class: Notification Function: Software/Firmware/Config: N/A ICN Hard Disk Health Diagnostic failed. icn1 - No disks to process->

2223859 The TPS Reset Failed because of a VRE Reset Failure See long description. 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Extended Error: This indicates that a failure to reset the VRE properly caused the TPS Reset to fail. Please refer to the VRE Troubleshooting Guide for further troubleshooting this issue, which can be found in the MR Service Documentation.

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.