GE Healthcare · MR2249391112677898313747222362122236252249420
TPS download problem Due to the network switch in the System Cabinet.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2249391 semTake failed for semaphore ID 0x3FFFDF0 1126778983 2 1 Thu Sep 15 10:09:43 2005 2249391 4 MR3T diagCoord sysIntegrity.c 1010 Task ID: 0x3fffbb0 Time: 13747 Object: APS semTake failed for semaphore ID 0x3FFFDF0. Errno = 0x3D0004. EN 218 SR 218 2223621 Internal Software Error. Can''t callback on undefined NMRID:HOST_SERVER. Tue Jun 15 09:05:04 2004 Error: 2223621 Host: IRIS Process: scn File: mrmail_connect_mgr.c Line: 232 Internal Software Error. Can''t callback on undefined NMRID:HOST_SERVER. 2223625 Send packet opcode:9004 seqnum:13 rev:2 type:1 length:16 req:SCAN_01_NOR resp:HOST_SERVER status:0 failed. Tue Jun 15 09:05:04 2004 Error: 2223625 Host: IRIS Process: scn File: mrmailmgr.c Line: 746 Send packet opcode:9004 seqnum:13 rev:2 type:1 length:16 req:SCAN_01_NOR r esp:HOST_SERVER status:0 failed. 2249420 A Monarch critical error was detected in task defaultServer Error: 2249420 Host: GEMR3T Process: diagCoord File: excLog.c Line: 1714 Task ID: 0x6eca750 Time: 3465 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A A Monarch critical error was detected in task defaultServer (0x3A94E20). Additional information in /usr/g/service/log/mxExLg-d
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
- Component
- 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- Images failed to reconstruct due to a failed Image Compute Node.
- MGD downlaod fails due to defective inifiniband board.
- Internal Software Errors due to Network Switch.
- CWA ID117781. Cannot scan with HOST Communications errors.
- CWA ID121877. TPS not responding during applications boot with Error 2223817.