GE Healthcare · MRExciteSigna22238780016311194621391225802922196282223625
Images failed to reconstruct due to a failed Image Compute Node.
Symptom
Lost images on Vibrant series. Customer is not sure if scan continues through completion or stops early. They will only get approximately 100 of 400+ expected images. Some images are also missing in the first series of 100 images ( not sequential). Customer performs successful TPS reset and continues with scheduled patients without any further problems. Itrak# 13163660
Error message
2223878 A process has crashed on the VRE (00 30 48 59 F0 64[icn1]). Process dpserver, signal causing the crash 7, time 001631, date: 2007-11-10, UTC 1194621391 2223878 HDe_mfg NSP VREReset.cc 1008 Exception Class: Unknown Function: Recon: Data Acquisition A process has crashed on the VRE (00:30:48:59:F0:64[icn1]). Process: dpserver, signal causing the crash: 7, time: 00:16:31, date: 2007-11-10, UTC: 1194621391. Extended Error: During a TPS Reset the VRE gets scanned for core files that were generated since the last TPS Reset or System Restart. Only one such message is generated even if there is more than once core file. The oldest core file found across all ICNs is listed in this message. 2258029 Data mover error: Most likely recon is not keeping up with incoming data rate. Error: 2258029 Host: OPICMR3T Process: LxMxConsumer File: FrameRcvHWO.cpp Line: 466 Server Name: DP ^C Data mover error: Most likely recon is not keeping up with incoming data rate. Please check traces. 2219628 The BAM Check Packet failed to reach the TPS Tue Jun 15 08:58:55 2004 Error: 2219628 Host: IRIS Process: scn File: SrxIPScreen.m Line: 1956 The BAM Check Packet failed to reach the TPS. 2223625 Send pa
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.